Vanadis Core T21/T18/T13 Reagent Cartridge
Brand(s)
Last updated
Summary
Product
Vanadis Core T21/T18/T13 Reagent Cartridge
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Vanadis Core T21/T18/T13 Reagent Cartridge | 746100 745859 745619 744592 742254 745232 746522 741096 744481 744811 746077 743926 745136 744491 | 3222-001B_CA |
Issue
Affected lots can include plates, which are slightly higher than normally, and the plates may rest higher on the instrument adapter (plates manufactured by our supplier using a different mold). This may result in run failure due to issues with moving the plate within the instrument or, due to suboptimal temperature during extraction of cell-free DNA, can cause insufficient protein degradation, which can potentially affect performance of the Vanadis Core® T21/T18/T13 reagent cartridge 3222-001B_CA.
Recall Start Date: August 3, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies
Wallac Oy
Mustionkatu 6, Turku, Finland, 20750
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74118