Stryker ProCuity Beds
Brand(s)
Last updated
Summary
Product
Stryker ProCuity Beds
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Stryker ProCuity Beds | More than 10 numbers, contact manufacturer. | 3009PX-LE-400 300900000000 3009QC-P1 3009PX-LEX-400 |
Issue
Stryker has identified that the manufacturing records for certain ProCuity beds are missing test values for electrical safety tests as required by IEC 62353:2014. As a result of the missing test data, the impacted units are not compliant to the above standard. Therefore, Stryker is initiating this voluntary recall to perform the necessary testing and documentation in the field.
Recall start date : April 29, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Stryker Medical
3800 E Centre Avenue, Portage, Michigan, United States, 49002
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75497