Servo-N
Brand(s)
Last updated
Summary
Product
Servo-N
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Servo-N | v. 4.4 or 4.5 with HFO-option | 6688600 |
Issue
A limitation of the intended performance of Servo-N during high frequency oscillatory ventilation (HFOV) when using the Fisher & Paykel Neonatal Circuit (950N81*) and the humidifier FP950. The use of a relatively small endotracheal (ET) tube in relation to body weight and in combination with the Fisher & Paykel Neonatal Circuit* that is used with the FP950 humidifier limits the delivered Pampl and therefore reduces the delivered VThf.
Recall Start Date: January 22, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Maquet Critical Care Ab
Rontgenvagen 2, Solna, Sweden, 17154
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74999
Get notified
Receive notifications for new and updated recalls and alerts by category.