Progressa Bed Surfaces
Brand(s)
Last updated
Summary
Product
Progressa Bed Surfaces
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Progressa Bed Surfaces | More than 10 numbers, contact manufacturer. | P7520X3 P7520X23 |
Issue
Baxter Corporation is issuing an urgent medical device correction for the Progressa Bed Surfaces (mattresses) listed to inform customers of a potential issue where the air bladders inside the mattress may move out of position when the head of the bed is elevated, causing a dip in the mattress.
Recall Start Date: February 25, 2025
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Physical medicine
Companies
Hill-Rom, Inc.
1069 State Route 46 East, Batesville, Indiana, United States, 47006-9167
Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Identification number
RA-77063