Hamilton-C6 Intensive Care Ventilator
Brand(s)
Summary
The local approved distribution partner is the first point of contact.
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
Hamilton-C6 Intensive Care Ventilator | Serial numbers from 13636 till 13950 Serial numbers from 13966 till 14753 | 160021 |
Issue
Hamilton Medical AG received information from the supplier of the ESM-board which is installed in the Hamilton-C6, indicating that certain ESM-board may contain a mechanically damaged ceramic capacitor. Since only certain serial numbers (from 13636 till 13950 and from 13966 till 14753) of the Hamilton-C6 are affected with the defective ESM-board, this FSN is limited to these devices defined above. Damage to the capacitor could, over time, lead to a loss of its function due to oxidation or similar degradation processes. The worst-case scenario is a local short circuit of the aforementioned capacitor, with different outcomes depending on the device state.
Recall start date: June 2, 2025
Additional information
Details
Hamilton Medical Ag |
Via Crusch 8, Bonaduz, Switzerland, 7402 |