Health product recall

Hamilton-C6 Intensive Care Ventilator

Last updated

Summary

Product
Hamilton-C6 Intensive Care Ventilator
Issue
Medical devices - Performance
What to do

The local approved distribution partner is the first point of contact.

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Hamilton-C6 Intensive Care Ventilator Serial numbers from 13636 till 13950 Serial numbers from 13966 till 14753 160021

Issue

Hamilton Medical AG received information from the supplier of the ESM-board which is installed in the Hamilton-C6, indicating that certain ESM-board may contain a mechanically damaged ceramic capacitor. Since only certain serial numbers (from 13636 till 13950 and from 13966 till 14753) of the Hamilton-C6 are affected with the defective ESM-board, this FSN is limited to these devices defined above. Damage to the capacitor could, over time, lead to a loss of its function due to oxidation or similar degradation processes. The worst-case scenario is a local short circuit of the aforementioned capacitor, with different outcomes depending on the device state.

Recall start date: June 2, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Hamilton Medical Ag
Via Crusch 8, Bonaduz, Switzerland, 7402
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77583

Get notified