DxC 500 AU Clinical Chemistry Analyzer
Brand(s)
Last updated
Summary
Product
DxC 500 AU Clinical Chemistry Analyzer
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
DxC 500 AU Clinical Chemistry Analyzer | 2024070118 | C63520 C63519 |
Issue
Beckman Coulter has determined that the system cannot recalculate a calculated result when results are rerun within the calculated test. The system will report only the calculated result using the initial constituent tests' results to the operator and lis/middleware. A patient may receive unnecessary treatment or testing; or patient treatment may be delayed.
Recall Start Date: February 13, 2025
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies
Beckman Coulter, Inc.
250 S. Kraemer Blvd., Brea, California, United States, 92821
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77067