Health product recall

BD Facsduet™ Sample Preparation Systems

Last updated

Summary

Product
BD Facsduet™ Sample Preparation Systems
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Bd Facsduet™ Sample Preparation System (Base) 9220010275 9220010313 662588
Bd Facsduet™ Premium Sample Preparation System 9230000104 9230000114 9230000112 9230000105 9230000138 9230000106 666339

Issue

Through a customer complaint investigation, BD identified a specimen volume bias issue with the BD Facsduet™ sample preparation system when using software version 1.4.1.
When the multidispense feature is enabled in the BD Facsduet™ sample preparation system software, and the user is preparing a two-tube assay, the specimen volume dispensed into the first tube will be at least 3% higher than the specimen volume dispensed into the second tube. For assays, where absolute counts are calculated using BD Trucount™ tubes, an increase in the volume may result in higher-than-expected absolute counts in the first tube compared to the second tube.
This issue will be observed when preparing the following assays on the BD Facsduet™ sample
preparation system:
1. BD Multitest™ IMK kit with BD Trucount™ tubes.
note: if using the BD Multitest™ IMK kit with BD Trucount™ tubes, a QC message will appear on your report if the bias is outside the acceptable range per assay requirements.
2. BD Multitest™ 6-color TBNK with BD Trucount™ tubes only if consecutive replicates of two are run.
3. for all other quantitative assays utilizing BD Trucount™ tubes in consecutive replicates of two, please follow your institution's procedures to determine if a review of reported results is necessary.
This issue does not affect assays that are determining percent positive, semi quantitative information.

Recall start date: August 12, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies
Becton, Dickinson And Company Bd Biosciences
155 North Mccarthy Boulevard, Milpitas, California, United States, 95035
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75970