Alinity M System
Brand(s)
Last updated
Summary
Product
Alinity M System
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Alinity M System | All lots. | 08N53-002 |
Issue
Abbott has identified an issue concerning the use of certain third-party universal transport medium (UTM) tubes on the Alinity M System. This issue has been observed with Copan® UTM-RT 3 mL without beads transport & preservation medium for viral molecular diagnostics testing, part number (PN) 3C047N. However, it is possible that other third party manufactured tubes could experience the same issue.
Recall Start Date: March 22, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies
Abbott Molecular Inc.
1300 E. Touhy Ave., Des Plaines, Illinois, United States, 60018
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75361
Get notified